IQI (Image Quality Indicator) Selection — ASME Section V Article 2 Complete Guide

IQI Selection — ASME Section V Article 2 Complete Guide | WeldFabWorld

IQI (Image Quality Indicator) Selection — ASME Section V Article 2 Complete Guide

The Image Quality Indicator — universally known in the field as the IQI, and also called a penetrameter — is the single most important quality control device in industrial radiographic testing. Every radiograph produced under ASME Section V Article 2 must demonstrate that the radiographic technique achieved the minimum required sensitivity, and the IQI is the tool that provides that demonstration. Get IQI selection or placement wrong, and every radiograph taken with that incorrect technique is invalid — requiring re-exposure, re-inspection, and the delays and costs that follow.

Despite its critical importance, IQI selection is one of the topics where radiographers and welding inspectors most frequently make errors in the field. The errors take several forms: selecting the wrong IQI number for the material thickness, placing the IQI on the wrong side of the weld, forgetting the F marker for film-side placements, failing to place IQIs at both ends of long weld exposures, and misidentifying the 2T hole requirement. These errors are not matters of academic principle — they have real consequences: rejected radiograph packages during ASME Code inspection, failed surveillance audits by the Authorised Inspector, and in the worst cases, pressure components in service that have not been adequately examined.

This article provides a complete, code-referenced guide to IQI selection and use under ASME Section V Article 2. Every rule is stated with its paragraph reference, every selection table is explained with practical worked examples, every common error is identified and corrected, and a 30-question timed quiz tests your knowledge across the complete subject.

Code Reference: All rules and paragraph references in this article are from ASME Section V, Article 2 — Radiographic Examination. The most commonly applied edition in the process industries is ASME Section V-2021, but always confirm against the edition specified in your Engineering Design Basis or project contract. Notable changes in IQI requirements have occurred between editions — particularly around wire IQI essential wire selection.

What is an IQI and Why is it Required? T-260

An Image Quality Indicator is a small reference device, made from a material radiographically similar to the object being examined, that is placed on the object during radiographic exposure. When the radiograph is processed, the image of the IQI appears on the film. The analyst examines this IQI image to verify that the radiographic technique produced an image of sufficient quality — that small details of specified size are clearly visible.

What the IQI Does NOT Do

This is the single most important conceptual point about IQIs, and it is frequently misunderstood by trainees and even experienced practitioners:

Critical Misunderstanding — IQI Sensitivity Is NOT Defect Detection Sensitivity: The IQI does NOT measure the smallest defect that will be detected in the weld. A 2% sensitivity (the 2-2T standard) does NOT mean the radiograph can detect a defect that is 2% of the wall thickness in size. The IQI measures only that the image quality is sufficient — that the film contrast, definition, and density are adequate. Actual defect detection depends on many additional factors: defect orientation relative to the beam, defect geometry (planar vs volumetric), beam divergence, scatter, and the radiographer’s interpretation skill. The IQI is a quality indicator for the imaging technique, not a guarantee of defect detection at the stated sensitivity level.

The IQI as Technique Validation

ASME Section V T-261 states the purpose clearly: IQIs are required to provide visual evidence that the radiographic examination technique produces the required radiographic sensitivity. If the required IQI image is not visible on the radiograph, the technique has not met the code minimum sensitivity requirement and the radiograph is invalid — the examination must be repeated with an improved technique.

Hole-type vs Wire-type IQI T-233, T-276, T-283

ASME Section V Article 2 recognises two distinct types of IQI, each with its own design standard, selection table, and sensitivity acceptance criterion. Both types are acceptable under the code; the choice is at the discretion of the examination procedure, provided the correct selection table is used for the chosen type.

Hole-type IQI (Plaque / ASTM type)
  • Design: Flat rectangular metal plaque of precisely controlled thickness, containing three drilled holes: 1T hole (diameter = IQI thickness), 2T hole (diameter = 2 × IQI thickness), and 4T hole (diameter = 4 × IQI thickness)
  • Identification: IQI number stamped on the plaque — designates thickness in thousandths of an inch (e.g. IQI No. 10 = 0.010 inch = 0.25 mm thick)
  • Selection table: T-276 — based on nominal single-wall radiographic thickness
  • Acceptance criterion: 2T hole visible (standard) — the hole whose diameter equals twice the IQI thickness
  • Standard: ASTM E1025 for manufacture
  • Common use: USA, ASME-governed projects worldwide
Wire-type IQI (DIN / EN / ISO type)
  • Design: Set of parallel metal wires of progressively increasing diameter (typically 6 or 7 wires per set) encapsulated in plastic. Wire diameters are standardised in a geometric progression.
  • Identification: Wire sets identified by set letter (A, B, C, D) and wire numbers 1–19. Lower numbers = smaller diameter wires.
  • Selection table: T-283 — specifies the essential wire number for each nominal radiographic thickness range
  • Acceptance criterion: Essential wire visible — the minimum wire diameter corresponding to the nominal radiographic thickness must be clearly visible on the radiograph
  • Standard: ASTM E747 (or EN ISO 19232-1) for manufacture
  • Common use: Europe, offshore projects, alongside or instead of hole-type IQIs
IQI Types — Hole-type and Wire-type HOLE-TYPE IQI (Plaque) 10 1 0 1T hole Ø = T 2T hole (REQUIRED visible) Ø = 2T ← KEY 4T hole Ø = 4T Width (varies by IQI number) T (thickness) IQI No. 10 = T = 0.010 in (0.254 mm) 2T hole diameter = 0.020 in (0.508 mm) WIRE-TYPE IQI Wire No. 10 11 12 13 Essential Wire 14 15 16 Increasing wire diameter → Essential wire = smallest wire clearly visible Must match or exceed T-283 requirement for wall thickness Correct Placement — IQI on Source Side of Weld IQI (source side) Radiation source above Film / Detector (below component)
Figure 1. Left: Hole-type IQI showing the three drilled holes (1T, 2T, 4T) with diameters equal to 1×, 2×, and 4× the IQI plaque thickness T respectively. The 2T hole (red circle) is the standard acceptance criterion. Right: Wire-type IQI showing parallel wires of increasing diameter — the essential wire (red) is the smallest wire that must be visible on the radiograph. Bottom: Both IQI types are placed on the source side of the weld (between the radiation source and the component), directly on the weld adjacent to the area being examined.

IQI Selection — Tables T-276 and T-283 T-276, T-283

IQI selection is a code-mandated process, not an engineering judgement call. The applicable ASME Section V table specifies exactly which IQI designation number (for hole-type) or essential wire number (for wire-type) must be used for each range of nominal single-wall radiographic thickness. Using a thinner IQI than required produces an artificially good sensitivity result — the IQI appears on the film more easily than warranted. Using a thicker IQI than required is overly conservative and will cause unnecessarily frequent failed technique validations.

Hole-type IQI — Table T-276 (Representative Values)

Nominal Single-Wall Radiographic Thickness IQI Designation (No.) IQI Thickness (in) IQI Thickness (mm) 2T Hole Diameter (mm)
Up to 6.4 mm (1/4 in)50.005 in0.127 mm0.254 mm
6.4 to 9.5 mm (1/4 to 3/8 in)70.007 in0.178 mm0.356 mm
9.5 to 12.7 mm (3/8 to 1/2 in)100.010 in0.254 mm0.508 mm
12.7 to 19.0 mm (1/2 to 3/4 in)120.012 in0.305 mm0.610 mm
19.0 to 25.4 mm (3/4 to 1 in)150.015 in0.381 mm0.762 mm
25.4 to 50.8 mm (1 to 2 in)200.020 in0.508 mm1.016 mm
50.8 to 76.2 mm (2 to 3 in)250.025 in0.635 mm1.270 mm
76.2 to 101.6 mm (3 to 4 in)300.030 in0.762 mm1.524 mm
101.6 to 127.0 mm (4 to 5 in)350.035 in0.889 mm1.778 mm
127.0 to 152.4 mm (5 to 6 in)400.040 in1.016 mm2.032 mm
Note on Table T-276: These values are representative of the standard ASME Section V table. Always use the actual code table for the applicable edition — intermediate IQI numbers (6, 8, 10, 12, 15, 17, 20, 25, 30, etc.) exist in the full table for the complete thickness range. The IQI number directly equals the IQI thickness in thousandths of an inch. IQI No. 20 = 0.020 inch thick. The 2T hole diameter for IQI No. 20 = 2 × 0.020 = 0.040 inch = 1.016 mm.

Wire-type IQI — Table T-283 (Representative Values)

Nominal Single-Wall Radiographic Thickness Essential Wire Number (ASTM E747) Essential Wire Diameter (in) Essential Wire Diameter (mm) Approx. Sensitivity (%)
Up to 6.4 mm (1/4 in)40.008 in0.203 mm~3.2%
6.4 to 9.5 mm50.010 in0.254 mm~2.7–4%
9.5 to 12.7 mm60.013 in0.330 mm~2.6–3.5%
12.7 to 25.4 mm70.016 in0.406 mm~1.6–3.2%
25.4 to 50.8 mm90.025 in0.635 mm~1.25–2.5%
50.8 to 101.6 mm110.040 in1.016 mm~1.0–2.0%
101.6 to 152.4 mm130.063 in1.600 mm~1.05–1.57%
Wire IQI — Higher Wire Number = Larger Diameter Wire = Thicker Material: Wire numbers in ASTM E747 are counterintuitive to some users: a higher wire number means a larger wire diameter. For thicker material requiring deeper penetration, a larger-diameter (higher-numbered) essential wire is required. Wire 4 (0.008 in diameter) is for thin material; Wire 13 (0.063 in diameter) is for thick material. When you read “essential wire = Wire 9”, it means wire number 9 with its specific diameter must be the smallest wire clearly visible on the radiograph.

Radiographic Sensitivity and the 2-2T Rule T-274

The concept of radiographic sensitivity is expressed in terms of the IQI used. The standard ASME Section V sensitivity expression is stated as a ratio: the IQI designation number as a percentage of the nominal radiographic thickness, followed by the hole that must be visible.

Standard ASME Sensitivity Expression (Hole-type IQI):
Sensitivity = (IQI thickness / Nominal radiographic thickness) × 100%

Standard requirement: 2% sensitivity with 2T hole visible = “2-2T”
This means:
IQI thickness = 2% of nominal single-wall thickness
AND: the 2T hole (diameter = 2 × IQI thickness = 4% of wall) must be visible

Example — 25.4 mm (1 inch) wall, IQI No. 20:
IQI No. 20 = 0.020 in (0.508 mm) = 2% of 1.000 inch → meets 2% requirement
2T hole = 0.040 in (1.016 mm) diameter → must be visible on film
If 2T hole is visible: technique achieves 2-2T sensitivity. Radiograph acceptable.
If 2T hole is NOT visible: technique is inadequate. Re-expose with improved technique.

Better sensitivity levels (achieved by improved technique):
1-1T: IQI No. 10 (1%) with 1T hole visible → better than required
2-1T: IQI No. 20 (2%) with 1T hole visible → higher than standard

The “2-2T” designation is the standard minimum sensitivity requirement for ASME pressure vessel and pressure piping radiography. Better sensitivity than 2-2T is always acceptable and preferred — it means the film quality is better than the minimum required. The code only establishes a minimum floor; better technique that achieves 1-2T or 2-1T sensitivity is entirely acceptable and may be required by some Owner specifications for critical weld examinations.

IQI Placement Rules T-277

ASME Section V T-277 specifies exactly where the IQI must be placed on the part. The placement rules are designed to ensure that the IQI image on the film represents the worst-case image quality that the technique can achieve over the weld area being examined — if the IQI is visible in the worst position, it will also be visible in all better positions within that exposure.

T-277.1 — General Placement Rules

  • The IQI shall be placed on the source side of the part being examined — between the radiation source and the part. This is the preferred standard position.
  • The IQI shall be placed on the weld or adjacent to the weld within the area of interest being radiographed. The IQI must be within the area of diagnostic interest shown on the radiograph.
  • The IQI shall be placed with the body of the IQI (for hole-type) or the wire set (for wire-type) within the area being examined — not on the edge of the film holder or outside the area of interest.
  • The lead identification numbers and letters on the IQI must be visible on the radiograph and must not obscure the area being examined for discontinuities.
  • For hole-type IQIs, the IQI identification number must be on the end of the IQI away from the weld, so the IQI numbers do not overlap with the weld image.

T-277.1(a) — IQI on Weld vs Adjacent Material

The IQI may be placed directly on the weld surface (if the weld surface condition permits consistent probe coupling) or adjacent to the weld on the base material at the same nominal thickness as the weld being examined. If the weld cap creates a significant thickness variation that would affect the IQI image, the IQI is typically placed on the base metal adjacent to the weld rather than on the weld cap.

T-277.2 — IQI Placement for Welds with Backing Strips

When a weld is made with a permanent backing strip (backing bar), the backing strip thickness must be included in the nominal radiographic thickness used for IQI selection. If the backing strip is present during radiographic examination, it adds to the total material thickness that the radiation must penetrate, which requires a thicker (higher-numbered) IQI than would be needed without the backing strip.

Source-side vs Film-side IQI Placement T-277.1(b)

Source-side placement is the standard and preferred position for all IQIs in ASME Section V radiographic examination. However, there are practical situations — particularly when radiographing the inside of vessels or pipe bends from outside — where access to the source side is restricted and the IQI must be placed on the film side. ASME Section V provides specific rules for both situations.

IQI Placement — Source-side (Standard) vs Film-side (Permitted) SOURCE-SIDE (Standard) Source IQI Pipe Wall Film Preferred position No F marker needed Standard sensitivity table applies FILM-SIDE (Permitted with F marker) Source Pipe Wall IQI + F Film F F marker MANDATORY One IQI designation number better required (per T-277.1(b) Note)
Figure 2. Source-side IQI placement (left, preferred) places the IQI between the radiation source and the pipe wall. Film-side placement (right, permitted when source-side access is impractical) places the IQI between the pipe wall and the film. Film-side placement requires a mandatory lead letter F marker adjacent to the IQI on the radiograph (ASME V T-277.1(b)) and requires that the IQI be one designation number better (thinner) than the source-side table requirement for the same wall thickness.

Film-side Placement Requirements T-277.1(b)

When source-side IQI placement is impractical (documented in the examination record), film-side placement is permitted under the following specific conditions:

  1. Lead letter F marker: A lead letter “F” must be placed adjacent to the IQI on the part, so that the letter F appears on the radiograph next to the IQI image. This permanently records on the film that a film-side IQI was used.
  2. One designation number better: For film-side placement, the IQI selected must be one designation number better (thinner) than the table requirement for the nominal radiographic thickness. If the source-side table requires IQI No. 20, film-side placement requires IQI No. 17 (the next thinner standard IQI).
  3. Justification documented: The examination record must document why source-side placement was impractical — it is not acceptable to use film-side placement simply for convenience.
Most Common IQI Non-conformance — Missing F Marker: The absence of the lead F marker on a radiograph taken with a film-side IQI is one of the most frequently cited IQI non-conformances during ASME Code surveillance audits. The film may show a perfectly visible IQI image meeting the sensitivity requirement, but without the F marker, the auditor cannot confirm whether the more stringent film-side requirement (one designation number better) was met or whether the standard source-side table was incorrectly applied. The F marker is not optional — it is mandatory per T-277.1(b) whenever a film-side IQI is used.

Number of IQIs Required per Exposure T-277.2

ASME Section V T-277.2 specifies the minimum number of IQIs that must be used based on the length of weld being radiographed in a single exposure:

Weld Length in Single Exposure Minimum IQIs Required IQI Placement Notes
Up to 250 mm (10 in) 1 IQI Near one end of the area being examined Standard single-IQI exposure; most individual pipe girth weld exposures
Greater than 250 mm (10 in) 2 IQIs One near each end of the area being examined Required for long weld exposures such as vessel seam welds and large-diameter pipe panoramic shots
Panoramic exposure — full circumference Minimum 3 IQIs At equally spaced intervals around the circumference (approximately 120 degrees apart) For panoramic single-wall exposures using an internal source; placement per T-277.2(b)
Multiple welds in a single exposure One IQI per weld On each weld or adjacent to each weld within the area of interest Each weld must be individually qualified even when radiographed simultaneously
Why Two IQIs for Long Exposures? For welds longer than 250 mm in a single exposure, film density and image quality at the ends of the exposure are likely to differ from those at the centre. X-ray intensity follows the inverse-square law, and the angle of incidence changes across long welds. By placing one IQI at each end of the exposed weld length, the radiographer verifies that the minimum sensitivity requirement is met across the entire exposed length — not just at one position. If the end IQIs are visible but the centre is overexposed or vice versa, the technique requires adjustment.

IQI Geometry — Dimensions in Detail T-233, ASTM E1025, E747

Hole-type IQI Dimensional Requirements

Per ASTM E1025 (referenced by ASME Section V), hole-type IQIs must meet precise dimensional requirements:

  • Thickness tolerance: ±5% of the designated thickness
  • Length and width: Minimum 25 mm (1 in) long and 10 mm (3/8 in) wide for most IQI numbers; larger IQIs have proportionally larger dimensions
  • Hole diameters: 1T hole, 2T hole, and 4T hole — each drilled perpendicular to the plaque face with diameter tolerance ±5%
  • Identification numbers: Lead numbers (or numbers of the same material as the IQI) that are 1T tall, permanently attached and visible on the radiograph
  • Material: Must be radiographically equivalent to or more difficult to penetrate than the material being examined — never use a more penetrable (lighter) material IQI on a denser workpiece

IQI Material Groups T-276.1

ASME Section V Table T-276.1 assigns materials to IQI material groups based on radiographic equivalence. The IQI material must be from the same group as or a more absorbing group than the material being examined:

IQI Material Group Materials Included Notes
Group 1 — Aluminium Aluminium and aluminium alloys Lightest group; use aluminium IQI for aluminium components
Group 2 — Titanium / Zirconium Titanium, titanium alloys, zirconium Intermediate group; use Ti or Zr IQI
Group 3 — Carbon / Low-alloy Steel Carbon steel, low-alloy steel (P-No. 1–5), cast iron Most common group for pressure vessel and piping radiography; steel IQI used
Group 4 — Stainless / Nickel Alloys Austenitic SS, nickel alloys, inconel, monel, duplex SS Steel IQI acceptable — steel is more absorbing than austenitic SS and provides more stringent test
Group 5 — Copper alloys Copper, copper-nickel, brass, bronze Copper or steel IQI required

Weld Reinforcement and IQI Selection T-274

The nominal radiographic thickness used for IQI selection is the single-wall thickness of the material being examined. ASME Section V T-274.1 specifies how weld reinforcement affects the IQI selection calculation:

Nominal Radiographic Thickness for IQI Selection (T-274.1):
For welds without reinforcement: t_IQI = single-wall base metal thickness
For welds with reinforcement left in place during RT: t_IQI = base metal thickness + weld reinforcement

Example: 25 mm wall pipe, weld cap reinforcement 3 mm (measured or calculated)
t_IQI = 25 + 3 = 28 mm → select IQI for 25.4 to 50.8 mm range → IQI No. 20
vs. without reinforcement: 25 mm → 19.0 to 25.4 mm range → IQI No. 15

Important: Weld reinforcement is included in t_IQI only if it is present during the RT examination.
If the weld cap is ground flush before RT, use only the base metal thickness for IQI selection.
Backing Strips: When a permanent backing strip is present and the radiograph includes the backing strip thickness (as is typical for most backing-strip weld configurations), the backing strip thickness must also be added to the nominal radiographic thickness for IQI selection. Failure to account for backing strip thickness results in use of an IQI that is too thin for the actual total radiographic thickness — producing a sensitivity that appears to be met on the film but is actually marginal given the total thickness being examined.

Special Cases and Exceptions

Radiographic Shims T-277.3

When the IQI is placed on a curved surface (such as a pipe curvature) and there is a gap between the IQI and the weld surface that would create a step change in radiographic thickness, a radiographic shim may be placed under the IQI to bring it to the same radiographic level as the weld being examined. The shim material must have the same radiographic equivalence as the weld material and must be sized to ensure the IQI sits at the correct radiographic level.

Single-Film Technique for Double-Wall Exposures

For double-wall radiographic techniques (where the radiation passes through both walls of a pipe and only one wall is being examined on the film), the IQI is placed on the source-side wall adjacent to the weld being examined. The nominal radiographic thickness for IQI selection is based only on the single wall containing the weld being evaluated — not the total double-wall thickness the radiation must penetrate.

Inaccessible Welds — Substrate Technique

When the weld being radiographed is in a location where the IQI cannot physically be placed on the weld itself (such as inside a closed vessel with no internal access), a separate substrate block of the same nominal radiographic thickness as the weld is radiographed simultaneously with the production exposure. The IQI is placed on this substrate, and the IQI image from the substrate exposure demonstrates the technique sensitivity. The examination record must document this alternative placement and reference the substrate qualification radiograph.

Acceptance Criteria for IQI Images T-274.2, T-285

A radiograph meets the IQI sensitivity requirement when the following conditions are all simultaneously satisfied:

  1. The IQI designation number is clearly visible on the radiograph and matches the required designation from Table T-276 (hole-type) or the required essential wire from Table T-283 (wire-type)
  2. For hole-type IQIs: the 2T hole is clearly visible (standard 2-2T requirement) — both its outline and its density difference from the background must be discernible
  3. For wire-type IQIs: the essential wire is clearly visible — its complete image across the width of the IQI must be distinguishable from the background
  4. The film density in the area of interest is within the specified range (typically 1.8 to 4.0 for X-ray film, or as specified in the applicable code edition)
  5. For film-side placement: the lead F marker is visible adjacent to the IQI image
  6. The IQI image does not overlap the weld or any area being examined for discontinuities (IQI identification numbers must be clear of the examination area)
What “Clearly Visible” Means: ASME Section V does not define a specific optical density difference that constitutes “visible.” The standard interpretation is that the feature (hole or wire) can be distinguished from the background by a qualified Level II or III RT interpreter under standard viewing conditions (densitometer-verified film density, appropriate viewing illuminator with adequate brightness, no specular reflections, magnifying glass if needed). If there is ambiguity about whether the 2T hole or essential wire is truly visible, the benefit of the doubt goes to the more conservative interpretation — re-expose with improved technique.

Common Errors in IQI Selection and Use

Error Consequence Correct Practice
Using wrong IQI number for the thickness Too thin IQI: technique appears to pass sensitivity when it does not. Too thick IQI: technique appears to fail when it actually passes. Always read directly from Table T-276 (hole) or T-283 (wire) using the actual nominal single-wall radiographic thickness including reinforcement if applicable
Film-side IQI without F marker Radiograph is non-conforming per T-277.1(b) regardless of IQI image quality; rejected at AI review Always place lead letter F adjacent to IQI before exposure when film-side placement is used
Placing IQI on the film side without applying the one-designation-number-better rule Apparent compliance — IQI visible — but technique sensitivity is actually lower than required for source-side equivalent When using film-side IQI, select one designation number thinner than the Table T-276 requirement for that thickness
Using only one IQI for a weld longer than 250 mm (10 in) Film density and sensitivity at the unmonitored end may be outside acceptable limits without detection Two IQIs — one near each end — for any single exposure covering more than 250 mm of weld
IQI placed outside the area of interest (beyond the edge of the diagnostic film) IQI image may not appear on the film at all, or may not represent the technique quality over the weld IQI must be within the area of interest on the film — verify IQI image is on film before accepting
Not including weld reinforcement in nominal radiographic thickness for IQI selection IQI selected is too thin — apparent sensitivity is better than the technique can actually achieve through the full weld crown Measure or calculate weld reinforcement and add to base wall thickness when selecting IQI
Using wrong material group IQI (e.g. aluminium IQI on steel) IQI appears too visible on film — does not represent true sensitivity through the denser steel Use IQI from the same group as or a more absorbing group than the material being examined (Table T-276.1)
Confusing 1T hole with 2T hole — claiming sensitivity when only the 1T hole was visible Over-claiming sensitivity: only the 2T hole is the standard acceptance criterion, not the 1T hole Know the dimensional relationship: 2T hole diameter = 2 × IQI thickness; verify which hole is visible by measurement

IQI Material Groups T-276.1

The IQI must be made from a material radiographically similar to or more dense than the material being examined. ASME Section V Table T-276.1 groups materials by radiographic equivalence. A steel IQI (Group 3) is acceptable for use on austenitic stainless steel (Group 4) because steel is more absorbing than austenitic stainless at typical X-ray energies — if the IQI is visible on the stainless steel radiograph, the sensitivity through the denser IQI material confirms adequate technique. The reverse is not permitted: a lighter material IQI (aluminium) cannot be used on steel because the aluminium IQI would appear too readily on the steel radiograph, creating a false impression of good sensitivity that the technique has not actually achieved through the steel material.

Recommended References and Study Materials

ASME Section V — Nondestructive Examination
The code volume containing Article 2 (radiographic examination) with all IQI selection tables, placement rules, and acceptance criteria. Essential reference for every RT practitioner and welding inspector.
View on Amazon
Radiography in Modern Industry — Kodak
Classic comprehensive reference for industrial radiography — covers IQI selection, film selection, exposure calculations, processing, interpretation, and acceptance criteria with detailed worked examples.
View on Amazon
ASNT NDT Handbook — Radiographic Testing
The ASNT comprehensive reference for RT — covers all aspects of industrial radiographic testing including IQI theory, selection, placement, and code compliance across major industries and standards.
View on Amazon
Industrial Radiography — Image Forming Techniques
Technical reference covering radiographic image quality, IQI sensitivity theory, geometric unsharpness, film processing, and digital radiography — bridges theory to field practice for RT Level II/III personnel.
View on Amazon

Disclosure: WeldFabWorld participates in the Amazon Associates programme (StoreID: neha0fe8-21). If you purchase through these links, we may earn a small commission at no extra cost to you. This helps support free technical content on this site.

Frequently Asked Questions

What is an IQI and what is its purpose in radiographic examination?
An Image Quality Indicator (IQI), also called a penetrameter, is a reference device placed on the object being radiographed to verify that the radiographic technique has sufficient sensitivity. The IQI does not measure defect detectability directly — it demonstrates that the film, exposure, and geometry combination can produce images of specified small details. If the required IQI image is visible, the technique meets the minimum code sensitivity. ASME Section V T-261 requires IQIs for all radiographic examinations of welds in pressure-retaining components.
What are the two types of IQI recognised by ASME Section V Article 2?
ASME Section V recognises two IQI types: (1) Hole-type IQI (ASTM type) — a flat metal plaque with three drilled holes (1T, 2T, 4T) identified by number designating thickness in thousandths of an inch; selected from Table T-276; 2T hole must be visible for acceptance. (2) Wire-type IQI — parallel metal wires of progressively increasing diameter; selected from Table T-283; the essential wire must be visible. Both are acceptable; the procedure specifies which type is used.
What is the 2T hole rule for hole-type IQI acceptance under ASME Section V?
The 2T hole rule is the standard ASME acceptance criterion: the radiograph is acceptable when the 2T hole — the hole whose diameter equals twice the IQI thickness — is clearly visible on the radiograph along with the IQI designation number. The 2T hole diameter equals 2 × IQI thickness. For example, IQI No. 20 (0.020 inch thick) has a 2T hole of 0.040 inch diameter. If the 2T hole is not visible, the radiographic technique is inadequate and must be improved — improved film type, increased exposure time, reduced scatter, or geometry adjustment — until the 2T hole becomes clearly visible.
How is the correct IQI number selected for a given material thickness?
The IQI number is selected from ASME Section V Table T-276 based on the nominal single-wall radiographic thickness, which includes base metal wall thickness plus weld reinforcement if present during RT, and backing strip thickness if applicable. The IQI thickness is approximately 2% of the nominal radiographic thickness — this is the basis for the 2% sensitivity standard. For wire-type IQIs, the essential wire is selected from Table T-283 with the essential wire diameter also approximately 2% of nominal thickness. Always use the actual code table for the applicable edition — never interpolate or estimate.
What is the difference between source-side and film-side IQI placement?
Source-side placement (between radiation source and part) is the standard preferred position — it represents the most demanding geometric condition for the IQI image. Film-side placement (between part and film) is permitted when source-side access is impractical. Film-side placement requires: (1) a mandatory lead letter F marker on the radiograph per T-277.1(b); (2) an IQI one designation number better (thinner) than the Table T-276 source-side requirement; and (3) documented justification for why source-side placement was impractical. Without the F marker, the film does not comply regardless of IQI image quality.
How many IQIs are required for a single radiographic exposure?
Under ASME Section V T-277.2: one IQI is required for weld lengths up to 250 mm (10 inches) in a single exposure, placed near one end of the area being examined. Two IQIs are required for weld lengths greater than 250 mm — one near each end. For panoramic exposures covering the full circumference, a minimum of three IQIs placed at equally spaced intervals (approximately 120 degrees apart) are required. For multiple welds radiographed simultaneously, one IQI per weld is required.
What does IQI sensitivity mean in radiography, and what is the standard requirement?
Radiographic sensitivity expressed via IQI means the ratio of IQI thickness to nominal radiographic thickness, expressed as a percentage, with the hole or wire that must be visible. The standard ASME requirement is 2-2T: the IQI thickness equals 2% of nominal radiographic thickness, and the 2T hole (diameter = 4% of wall thickness) must be visible. Sensitivity does not mean the smallest detectable defect is 2% of wall thickness — it means the image quality is sufficient to record details of that size if they were present. Better sensitivity (lower percentage, smaller hole visible) is always acceptable.
What is the purpose of the lead letter F marker used with film-side IQIs?
The lead F marker permanently records on the radiograph that a film-side IQI placement was used. This allows any reviewer — including the Authorised Inspector and future auditors — to confirm that the more stringent film-side sensitivity requirement (one designation number better than source-side) was applied. Without the F marker, there is no way to determine from the radiograph alone whether the source-side or film-side table was used, making compliance verification impossible. The F marker is mandatory per ASME V T-277.1(b) and its absence is a code non-conformance regardless of IQI image quality.

Related Technical Guides

IQI Selection Quiz

30 questions — 20 seconds each — ASME Section V Article 2 Level

30Questions
20sPer Q
📷

IQI & Radiography Quiz

30 questions covering IQI purpose, hole-type and wire-type selection, 2T hole rule, placement rules, F marker, sensitivity, film-side placement, number of IQIs, and common errors. For RT Level II/III practitioners and welding inspectors.

20-second timer
Auto-advances on timeout
30 questions
All Article 2 IQI topics
Instant feedback
Code-referenced explanation after every answer
🏆
Graded result
Score, grade, and performance summary
0%
0/30

0
Correct
0
Wrong
0
Timed out